GDRVPWID
Pulse Width
Enters the specified value as the pulse width.
Syntax
GDRVPWI
D
/
.._._.:
Item Description/Default
Raw-
Number
Any real or integer number. Default unit is
ps.
Range is from the
sweep delay
(GDRVSWDE) plus the
sweep time to the
Equivalent
Softkey:
E#m
WIDTH
.
Option Required: Option 105. Option 101 is recommended.
Initial Value: 0.
Related Commands: GDRVCLPAR, GDRVRBW.
Example
OUTPUT
718;"MOV
GDRVPWID,lMS;"
Sets the pulse width to 1
ms.
Description
Unlike
ElQTER
WIDTH , GDRVPWID does not make the marker function active or display
the
Pulse
Param
softkeys.
Once you enter a value into GDRVPWID, that value is retained until you change it, or execute
DISPOSE ALL. Pressing
CPRESET)
or turning the spectrum analyzer off does not change the
value of GDRVPWID.
You can execute the GDRVPWID command two different ways. You can either execute the
GDRVPWID command directly (for example,
"GDRVPWID
1MS
;
‘I)
or use
the
MOV command
to move the value for the pulse width into the GDRVPWID command (for example,
"MOV
GDRVPWID,
1MS;
I’).
If you use the MOV command, no text is displayed in the active function
area during command execution.
The GDRVPWID command is a gate utility function (see “GDRVUTIL” for more information
about the gate utility functions). Executing a gate utility command without Option 105
installed in the spectrum analyzer generates the
COMMAND ERROR:
message and an SRQ 140.