Agilent Technologies 8590 TV Converter Box User Manual


 
SEGDEL Segment Delete
Example 2
With the addition of line 60,
the
upper and lower limit-line tables are no longer treated as
separate tables, but as one table. The segment is deleted from the upper and lower limit-line
tables (for the given frequency).
10 OUTPUT
718;"LIMIDEL;"
20 OUTPUT
718;"LIMIMODE
UPPER;"
25 OUTPUT 718;"LIMIFT
FREQ;"
30 OUTPUT
718;"LIMISEG
300MHZ,-30DB,FLAT;"
40 OUTPUT
718;"LIMIMODE
LOWER;"
50 OUTPUT
718;"LIMISEG
SOOMHZ
,-70DB,FLAT;"
60 OUTPUT
718;"LIMIMODE
UPLOW;"
70 OUTPUT
718;"SEGDEL
1;"
80
END
Description
Deletes
the current limit-line table.
Specifies the upper limit-line table.
Limit
lines
to be
based
on frequency.
Enters segment into the upper limit-
line
table.
Spec@es
the
lower
limit-line
table.
Enters a segment into the
lower
limit-
line table
Spe&@
both the upper and
lower
limit-line tables.
Deletes the segment from the upper
and lower limit-line tables.
The result of SEGDEL depends on the setting of the LIMIMODE command as shown in the
following table.
LIMIMODE Setting Result of SEGDEL
LIMIMODE UPPER Deletes
specified
segment from the upper limit-line table.
LIMIMODE LOWER Deletes
specilled
segment from the lower limit-line table.
LIMIMODE UPLOW Deletes specified segment from the upper and lower limit-line table.
LIMIMODE DELTA Deletes specified segment from the mid and delta limit-line table.
You may want to query LIMIMODE before using SEGDEL if you are unsure of the LIMIMODE
setting.
‘Ib
determine the number of each segment, you can use the softkeys accessed by
Edi%
Limit
to display the limit-line table. (Limit-line entries are sorted according to frequency or time.)
5.484