Emerson Process Management GmbH & Co. OHG 4-23
X-STREAM XE
Instruction Manual
HASXEE-IM-HS
04/2010
4
Installation
4.5.2 Installation - X-STREAM XE Field Housings
The number of actually available signal out-
puts, and also the number of built-in modules
with screw-type terminals, varies according to
the conguration of the unit.
A maximum of three modules with 36 termi-
nals each can be tted.
The terminals can be accessed by opening
the front panel of the instrument.
ELECTRICAL SHOCK HAZARD
Verify the power supply at installation site meets the specication given on
the analyzer´s nameplate label, before installing the instrument!
Verify power cables are disconnected and/or instrument is de-energized
prior to working at the terminals!
Verify the power cord is layed with a distance of at least 1 cm (0.4 in) to any
signal cable to ensure proper insulation from signal circuits!
Characteristics of terminals:
Accepted wire gauge:
0.14 ... 1.5 mm
2
(AWG 26 ... AWG 16),
end sleeves not required
Skinning length: 5 mm (0.2")
Thread: M2
Min. tightening torque: 0.25 Nm (2.21 in.lb)