Emerson Process Management GmbH & Co. OHG 4-23
X-STREAM XE
Instruction Manual
HASXEE-IM-HS
04/2010
4
Installation
4.5.2 Installation - X-STREAM XE Field Housings
The number of actually available signal out-
puts, and also the number of built-in modules
with screw-type terminals, varies according to
the conguration of the unit.
A maximum of three modules with 36 termi-
nals each can be tted.
The terminals can be accessed by opening
the front panel of the instrument.
ELECTRICAL SHOCK HAZARD
Verify the power supply at installation site meets the specication given on
the analyzer´s nameplate label, before installing the instrument!
Verify power cables are disconnected and/or instrument is de-energized
prior to working at the terminals!
Verify the power cord is layed with a distance of at least 1 cm (0.4 in) to any
signal cable to ensure proper insulation from signal circuits!
Characteristics of terminals:
Accepted wire gauge:
0.14 ... 1.5 mm
2
(AWG 26 ... AWG 16),
end sleeves not required
Skinning length: 5 mm (0.2")
Thread: M2