06/2010
Zero&Span Single
Cancel!
Start!
Zero Gas 0.000 ppm
Span Gas 500.000 ppm
Blowback!
Concentration 0.000 ppm
Status..
Press enter in this line to cancel any ongoing
calibration
Press enter in this line to start a zero & span
calibration procedure
Nominal zero gas concentration (can be set
in SETUP
6-24)
Nominal span gas concentration (can be set
in the SETUP menu
6-24)
Immediately start a blowback procedure
Currently measured gas concentration.
Open a submenu to see calibration status
summary for all channels
6-11
6.2.2.3.1 Zero & Span Single Menu
6.2.2 Control Menu
Start and control a valve supported calibrati-
on procedure for zero and span for a single
channel only.
Multi-channel unit:
In SELECT COMPONENT, select the chan-
nel to be calibrated.
Control..
Advanced Calibration..
Zero&Span Single..
Multi-channel unit:
Press left to enter SELECT COMPONENT,