6.2.3 Setup Menu
6.2.3.2.7 Setup Calibration Deviations
Setup..
Calibration..
Deviations..
Ch2
Deviations
ZeroDev 5.000 ppm
ZeroDev Total 7.000 ppm
Zero Deviation Reset!
SpanDev 100.000 ppm
SpanDev Total 230.000 ppm
Span Deviation Reset!
In the context of this menu, deviation means
the value, the current zero or span calibration
value is corrected by subsequent calibrations.
"ZeroDev." or "SpanDev." in concentration
units give the correction by the last performed
calibration.
"ZeroDev. total" or "SpanDev. total" in concen-
tration units give the total (sum of) corrections
of all the referred calibrations since the last
time, deviations have been reset.
Reset all zero or span deviations.
Note!
These functions are carried out immediately,
and there´s no undo!
Example:
A measuring channel shows zero drift of
10 ppm per week. It is calibrated once a
week.
After 3 weeks, DEVIATIONS would show:
ZeroDev: 10 ppm (= last calibr.)
ZeroDev Total: 30 ppm (= summary of 3
calibrations, carried out
within 3 weeks)
Multi-channel unit:
In SELECT COMPONENT select the chan-
nel to be setup.
Multi-channel unit:
Press
left to enter SELECT COMPONENT, to
change the settings for a different channel.