Emerson HASXEE-IM-HS TV Video Accessories User Manual


 
Emerson Process Management GmbH & Co. OHG6-44
X-STREAM XE
Instruction Manual
HASXEE-IM-HS
06/2010
Ch1
Ranges
Current Range: Range1
Range Start: 0.000 ppm
Range End: 50000.000 ppm
Starts & Ends..
Range Control: Manual
Auto Ranging..
6.2.3.3.1 Setup Measurement Ranges
6.2.3 Setup Menu
Specify the limits for up to 4 ranges per chan-
nel in one single menu, making it easier to
adjust several limits at a glance;
6-45.
Select the range to be congured.
Available options: Range1 ... Range4
Enter this menu to congure up to 4 ranges
per channel.
For the current range, specify start and end
concentration limits.
For the current range, specify how range
switching is done.
Available options:
Manual, Remote, Automatic
Note!
Selecting Remote or Automatic range con-
trol is not possible, if identical ranges end
values are specied (
6-45)!
Specify the switchover levels for up to 4
ranges per channel;
6-45.
Setup..
Measurement..
Ranges..