Emerson HASXEE-IM-HS TV Video Accessories User Manual


 
Emerson Process Management GmbH & Co. OHG 2-3
X-STREAM XE
Instruction Manual
HASXEE-IM-HS
04/2010
2
Technical Data
Interfaces, signal inputs and outputs
Interface signals are accessed in different ways depending on the analyzer model:
X-STREAM XEGC, XEGP:
standard: subminD plugs and sockets
optional: screw-type terminal adapters
X-STREAM XEF, XDF, XEFD: internal screw-type terminals
All versions provide a 2 RJ45 plugs for Ethernet connections as well as 1 USB and 1 mini USB
connector (eld housings internally only).
2.1 Common Technical Data
All models are supplied with
up to 5 analog outputs
(standard: 1 analog output
per channel)
electrical
specication
4 (0) - 20 mA (R
B
< 500 )
optically isolated from each other and from analy-
zer electronics
function
user-congurable activation and deactivation of
concentration levels
support for NAMUR NE 43 operation modes, con-
gurable via keypad and Modbus
4 relay outputs
electrical
specication
dry contacts
max. load. 30 V; 1 A; 30 W resistive
function
Each output can be congured to one of the fol-
lowing functions:
NAMUR NE 107 status signal
'Failure'
'Maintenance request'
'Out of specication'
'Function check'
1 of 2 concentration limits per channel,
Control signals for
external valve V1 ... V8,
external sample gas valve
external pump
zoom status for analog outputs
2 Modbus interfaces Ethernet (RJ45 sockets)
2 USB ports
specication USB 1.0
function
1 USB connector type A,
for connecting external storage devices
1 USB connector type mini AB,
for connecting external computers