Emerson HASXEE-IM-HS TV Video Accessories User Manual


 
Emerson Process Management GmbH & Co. OHG7-6
X-STREAM XE
Instruction Manual
HASXEE-IM-HS
10/2010
Before starting any calibration take care of
the table below, and the following sections,
describing general preparations for calibra-
tion procedures and how to perform such
calibrations!
7.4 Calibration Procedures
Type of Procedure
Menu Page
(CONTROL - )
Valves
Simulta-
neously
Calibrated
Channels
More Information
Manual calibration
ZERO CALIBRATION...
optional single channel page 7-17
SPAN CALIBRATION...
Advanced calibration
ADV.CALIBRATION - ZEROALL!
required all channels page 7-20ADV.CALIBRATION - SPANALL!
ADV.CALIBRATION - ZSCALALL
!
Remote calibration n.a. (via Modbus or Dig IN) recommended all channels page 7-32
Unattended calibration n.a. (via interval time) required all channels page 7-36
Furthermore you´ll find information about
calibration gases setup: 7-7.