Emerson HASXEE-IM-HS TV Video Accessories User Manual


 
Emerson Process Management GmbH & Co. OHG6-36
X-STREAM XE
Instruction Manual
HASXEE-IM-HS
06/2010
Ch2
Procedure
Purge Time: 10 s
Max. Time: 120 s
Zero Ranges: Together
Span Ranges Separately
Zero Method: Stability
Span Method: Instant
Test Mode: No
6.2.3.2.3 Setup Calibration Procedure
6.2.3 Setup Menu
If the system is setup accordingly, the access
code for level 3 must be entered to gain ac-
cess to this menu.
The maximum time to complete a calibration
procedure, if calibrated with stability method.
If not already regular nished, a calibration
will be terminated after this time.
Accepted values: 0 ... 600 seconds
Specify how to calibrate multiple ranges of a
channel.
Available options: Together, separately
Specify the calibration methods for zero and
span calibrations.
Available options: Stability, Instant
The time required to completely ll the gas line
with the related gas, when switching to zero
or span gas
1)
. Adjust according your system.
Accepted values:
0 ... [max. time, see next menu line]
Multi-channel unit:
In SELECT COMPONENT, select the chan-
nel to be congured.
Multi-channel unit:
Press
left to enter SELECT COMPONENT,
to change the settings for another channel.
Setup..
Calibration..
Procedure..
Set to Yes, to simply check, if the calibration is
still valid: now the instrument performs calibra-
tion procedures, without correcting the calibra-
tion parameters (simulation of calibrations).
Note!
Marked lines are available only if “Valves” in
INSTALLED OPTIONS is set to a value other
than none.
1)
see note on page 6-37