Emerson HASXEE-IM-HS TV Video Accessories User Manual


 
Emerson Process Management GmbH & Co. OHG 4-35
X-STREAM XE
Instruction Manual
HASXEE-IM-HS
04/2010
4
Installation
Frequently, several loads in one system are
controlled by several analyzer outputs, whe-
reby the power for the loads derives from a
common source.
Special care is needed when wiring the loads
to minimize interference from switching these
loads:
• avoid connecting the loads to their power
supply by a common line:
4.6.4 Driving Multiple Loads
Fig. 4-30: Common Line
• It is recommended the loads to be wired
separately, and each load connected se-
parately to the power supply. Beginning
at the distribution point, both the + and
the - wires for each load are laid together
to the load (Fig. 4-31). Interference is
further reduced if a twisted multi-core
cable is used.
Distribution point
Lay cables to the loads
together as far as possible
(using twisted-pair cables
if possible)
Fig. 4-31: Loads in Parallel
4.6 Installation - Notes on Wiring
J
L