Emerson HASXEE-IM-HS TV Video Accessories User Manual


 
Emerson Process Management GmbH & Co. OHG 4-33
X-STREAM XE
Instruction Manual
HASXEE-IM-HS
04/2010
4
Installation
Fig. 4-27: Shield Connector Terminal With Cable
4.6 Installation - Notes on Wiring
The shield connector must be ordered to t the cable diameter, and can be retrotted:
Ø 1.5 ... 6.5 mm (0.06" ... 0.25") part # ETC02019
Ø 5 ... 11 mm (0.2" ... 0.43") part # ETC02020
Ø 10 ... 17 mm (0.4" ... 0.66") part # ETC02021
Ø 16 ... 24 mm (0.63" ... 0.94") part # ETC02022
X-STREAM XEGP/XEGC with screw-type
terminal adapters
In order to avoid measured values being inu-
enced by external interference signals when
terminal adapters are in use, the signal cable
shieldings must be connected to the analy-
zer housing by means of shield connector
terminals:
Strip the signal cable to a length of 20 cm
(8"). Take care to not damage the braided
shield!
Pull up the contact part of the shield con-
nector terminal,
feed through the cable as illustrated in
Fig. 4-27, and
release the contact part down onto the
braided shield.
The result is a secure contact with the cable
shield, improving the unit’s immunity against
interference from other electronic devices.
Finally connect the individual wires as descri-
bed in section 4.5.1.