7.4.1 Preparing a Calibration
The next step is to assign the valves to the
channels: Continue with
7-15.
Example:
For our example we assume, that the analyzer
controls 4 internal and 3 external valves:
Internal valves are connected as shown
in g. 7-3.
3 external valves are labelled V5 ... V7,
and connected to digital outputs 5 .. 7
For this to be setup, enter the second page
of the Digital Outputs menu, and for each
output 5 ... 7
select System for the "... Node"
select the label of the connected valve
(V5, V6 or V7) for the "...Signal", as shown
in the lefthand gure.
Digital Outputs (X4.1) 1of3
Output5 Node: System
Output5 Signal: V5
Output6 Node: System
Output6 Signal: V6
Output7 Node: System
Output7 Signal:
V7
Verify which digital outputs are connected
to control your external valves, and how the
valves are labelled.
Next enter the menu page, showing these
outputs, and for each output select System in
the line "Outputn Node" (where "n" is replaced
by the output number).
Finally for each output setup the valve's label.