Software Menus
6.2.3 Setup Menu
6.2.3.4.3 Setup Digital Inputs
Setup..
In-/Outputs..
Digital Inputs..
Digital Inputs (X4.1) 1of3
Input1 Node: System
Input1 Function: ExtFail
Input1 Edge: Up
Input2 Node: Ch1
Input2 Function: Range1
Input2 Edge:
Down
If your instrument features optional digital
I/O boards, this menu appears, enabling to
congure the digital inputs.
Menu pages 1...3 (titled "Digital inputs (X4.1)")
enable to congure inputs 1...7 on the rst ex-
pansion board ('X4.1' in the menu title refers
to the instrument´s I/O connector X4.1).
For each input 1...7 specify within the "Node"
line the signal source.
Available options: System, Ch1 ... Ch5
(depending on the number of channels in-
stalled).
If any one of Ch1 ... Ch5 is selected, only
signals, valid for the selected channel are
selectable.
If System is selected, any system signal is
selectable.
Once the "Node" is specied, for each input
select the "Function" of that input (depending
on the node, the list of available signals vari-
es;
next page) and select, how the input
is to be triggered: by rising "Edge" (Up), or
falling "Edge" (Down).
Input3 Node:
System
Input3 Function: ExtFail
Input3 Edge: Up
Input4 Node: System
Input4 Function: ExtFail
Input4 Edge: Up
Input5 Node: Ch1
Input5 Function: Range1
Input5 Edge:
Down
Digital inputs (X4.1) 3of3
Input6 Node: System
Input6 Function: ExtFail
Input6 Edge: Up
Input7 Node: Up
Input7 Function: Ch3
Input7
Edge: