Emerson Process Management GmbH & Co. OHG6-58
X-STREAM XE
Instruction Manual
HASXEE-IM-HS
06/2010
6.2.3 Setup Menu
6.2.3.3.4.1.1 Setup Cross Interference Linearization Curve
Setup..
Measurement..
Cross Interference..
IF Source1 ... 4
Linearization Curve..
Reference value to normalize the linearization
curve.
Accepted range: -1E+9 ... +1E+9
Enter up to 4 polynomial factors to linearize
the interfering component´s input signal with
a fourth-order polynomial of the form
Ch1
IF Source1
Reference Value: 1
IF Polynomial a1: 1.00000000
IF Polynomial a2: 0.00000000
IF Polynomial a3: 0.00000000
IF Polynomial a4: 0.00000000
a4 x
4
+ a3 x
3
+ a2 x
2
+ a1 x
Note!
The menu gure to the left shows the default
setup for the polynomials, which relates to a
straight line.
So, if your IF source signal is linear, no fur-
ther actions or changes in this menu are
required.