Emerson HASXEE-IM-HS TV Video Accessories User Manual


 
Emerson Process Management GmbH & Co. OHG4-8
X-STREAM XE
Instruction Manual
HASXEE-IM-HS
04/2010
Plugs and sockets required for the electrical
connections are on the rear panel of the units
(g. 4-4 & 4-5).
X-STREAM XEGP analyzers provide an in-
ternal widerange power supply for worldwide
use.
X-STREAM XEGC analyzers are powered
by an external DC 24 V power supply unit,
optionally supplied with the unit. If a PSU
is not included in delivery, another unit can
4.5.1 X-STREAM XEGC, X-STREAM XEGP
4.5.1 Installation - X-STREAM XEGC, X-STREAM XEGP
Fig. 4-4: X-STREAM XEGC - Table Top Version Rear Panel
Gas inlets
Gas outlets
Power in
Fuse
Analog outputs / relay
outputs
USB connectors
RJ45 ethernet connectors
Serial interface
( RS 232 /
Modbus 485)
Digital in- /
outputs
Serial CAN
interface
Service interface
be used instead, provided it conforms to the
specications on 2-9.
X-STREAM XEGC / XEGP analyzers should
be operated in a horizontal position.
An optional kit with brackets and blindplate
enables to install XEGC models into a rack;
this is accomplished by means of four screws
(
2-7 ). The external PSU is optionally
available for rack mounting, too.
2 1 3
1: ME
2: + 24 V
3: 0 V ( )
HAZARD BY RACK
INSTALLATION
The optional brackets for
rack mounting ( 2-7) are
not designed to carry the
weight of the instrument!
Support the instrument, when
rack mounting!
Disregarding may cause per-
sonal injury and damaged
equipment.