Emerson HASXEE-IM-HS TV Video Accessories User Manual


 
Emerson Process Management GmbH & Co. OHG 5-13
X-STREAM XE
Instruction Manual
HASXEE-IM-HS
04/2010
5
Startup
5.7 Checking the Settings
Starting with the measurement display (
page 5-11), pressing any key except home will
access the MAIN MENU; from here, follow
these steps
:
(If the display is showing anything other than
the measurement display, press
home to return
to the measurement display rst).
5.7 Checking the Settings
Set the preferred language for the software;
each analyzer shipped with 3 out of below list
of available languages.
Currently available (may be extended by
future software versions.):
English, French, German, Italian, Spanish,
Polish, Portuguese
Note!
If you are unfamiliar with the language set:
5-12 shows the sequence of keys to set
a different language.
( )
If the system has been set up accordingly,
the code for access level 1 must be entered
at this point to enable access to the following
menu.
Note!
The factory setting for this unit is "no code
required”. For ease of operation, it is recom-
mended to use the factory settings for access
codes while setting up the unit for the rst
time. In the following sections, therefore, no
more reference will be made to any need for
entering a code.
The following sections are structured so that
the user can work through them one by one
after powering up the unit. After completing
these steps, the unit will be congured to the
user’s needs and properly functioning.
Setup
Display..
Alarm..
Measurements..
In-/Outputs..
Communication..
Alarms..
Installed Options..
Save-Load..
Display
Contrast..
Language: English
Phrase Version EN1.04B
Measurements..
Measurement Display..
Menu Access.. 10 Min
Auto Home: 10 Min