1.4 Comparison of Analyzer Models
1.4 Comparison of the Various X-STREAM Analyzer Models
1
2 19 in housing, table-top or rackmountable,
protection type: IP 20
1
1 19 in housing, table-top or rackmountable,
protection type: IP 20
Stainless steel wallmountable eld housing,
protection type: IP66 / NEMA 4X
Single or dual compartment design
Cast aluminium wallmountable eld housing,
protection type: IP66 / NEMA 4X
24V input with external power supply unit Internal wide range power supply unit Internal wide range power supply unit Internal wide range power supply unit
Max. 3 channels: 2 photometer + 1 WLD/O
2
max. 6 gas connections,
including 1 optional purge gas connection
Max. 4 channels in any combination
max. 8 gas connections,
1 optional extra connection for purge gas
Max. 4 channels in any combination
max. 8 gas connections,
1 optional extra connection for purge gas
Max. 4 channels in any combination
max. 8 gas connections,
including 2 optional purge gas connection
Options for gas lines: Flow sensor, pressure
sensor, infallible gas lines
Options for gas lines: Flow sensor, pressu-
re sensor, heating for physical components,
sample gas pump, 1 or 2 valve blocks, infalli-
ble gas lines
Options for gas lines: Flow sensor, pressure
sensor, heating for physical components,
sample gas pump, 1 or 2 valve blocks, infalli-
ble gas lines
Options for gas lines: Flow sensor, pressu-
re sensor, heating for physical components,
sample gas pump, 1 or 2 valve blocks, infalli-
ble gas lines
1-4 analog outputs, 4 relay outputs, 2 Modbus
Ethernet interfaces, 2 USB connectors
optional:
1 interface card with 7 digital inputs and
9 digital outputs
1 interface card with analog inputs
electrical interfaces accessible via sockets on
back of unit, optionally: screw-type terminal
adapters
(except for Ethernet & USB)
1-4 analog outputs, 4 relay outputs, 2 Modbus
Ethernet interfaces, 2 USB connectors
optional:
1 or 2 interface cards, each with 7 digital
inputs and 9 digital outputs
1 interface card with analog inputs
electrical interfaces accessible via sockets on
back of unit, optionally: screw-type terminal
adapters
(except for Ethernet & USB)
1-4 analog outputs, 4 relay outputs, 2 Modbus
Ethernet interfaces, 2 USB connectors
optional:
1 or 2 interface cards, each with 7 digital
inputs and 9 digital outputs
1 interface card with analog inputs
electrical interfaces on internal screw-type ter-
minal adapters
(except for Ethernet & USB)
1-4 analog outputs, 4 relay outputs, 2 Modbus
Ethernet interfaces, 2 USB connectors
optional:
1 or 2 interface cards, each with 7 digital
inputs and 9 digital outputs
1 interface card with analog inputs
electrical interfaces on internal screw-type ter-
minal adapters
(except for Ethernet & USB)
LCD LCD LCD, impact tested front panel LCD, impact tested front panel
Operating ambient temperature:
0 °C to +50 °C (32 °F to 122 °F)
Operating ambient temperature:
0 °C to +50 °C (32 °F to 122 °F)
Operating ambient temperature:
-20 °C to +50 °C (-4 °F to 122 °F)
Operating ambient temperature:
-20 °C to +50 °C (-4 °F to 122 °F)
Models available for use in explosive environ-
ments
Flameproof enclosure: approved for use in
hazardous areas (explosive environments)
Size: (DxHxW): max. ca. 440x130x220 mm
Weight: ca. 8 - 12 kg (17.6 - 26.5 lb)
Size: (DxHxW): max. ca. 411x133x482 mm
Weight: ca. 11 - 16 kg (24.3 - 35.3 lb)
Size: (DxHxW): ca. 265x400 (815)x550 mm
Weight: max. ca. 25 (45) kg / 55.1 (99.2) lb
Size: (DxHxW): max. ca. 222x512x578 mm
Weight: max. ca. 63 kg (138.9 lb)
For more detailed information: 1-15 For more detailed information: 1-17 For more detailed information: 1-19 For more detailed information: 1-24
X-STREAM XEGC
X-STREAM XEGP