Emerson HASXEE-IM-HS TV Video Accessories User Manual


 
Emerson Process Management GmbH & Co. OHG 9-43
X-STREAM XE
Instruction Manual
HASXEE-IM-HS
04/2010
9
Modbus Commands
9.4 List of Parameters and Registers - Sorted by Daniel Registers
Address
Tag Name
Data
Type
Client
Access
Description
Daniel Modicon
9026
6051…
6052
Setup.Measurement.EndOfCurren-
tRange_3
Float R/W end of current range of comp3
9027
6053…
6054
Setup.Measurement.StartOfCurren-
tRange_4
Float R/W start of current range of comp4
9028
6055…
6056
Setup.Measurement.EndOfCurren-
tRange_4
Float R/W end of current range of comp4
9029
6057…
6058
Setup.Measurement.StartOfCurren-
tRange_5
Float R/W start of current range of comp5
9030
6059…
6060
Setup.Measurement.EndOfCurren-
tRange_5
Float R/W end of current range of comp5
9031
6061…
6062
Status.Pressure1 Float RO pressure of comp1
9032
6063…
6064
Status.Pressure2 Float RO pressure of comp2
9033
6065…
6066
Status.Pressure3 Float RO pressure of comp3
9034
6067…
6068
Status.Pressure4 Float RO pressure of comp4
9035
6069…
6070
Status.Pressure5 Float RO pressure of comp5
9036
6071…
6072
Status.Flow1 Float RO ow of comp1
9037
6073…
6074
Status.Flow2 Float RO ow of comp2
9038
6075…
6076
Status.Flow3 Float RO ow of comp3
9039
6077…
6078
Status.Flow4 Float RO ow of comp4
9040
6079…
6080
Status.Flow5 Float RO ow of comp5
9041
6081…
6082
Status.Temperature1 Float RO temperature of comp1
9042
6083…
6084
Status.Temperature2 Float RO temperature of comp2
9043
6085…
6086
Status.Temperature3 Float RO temperature of comp3
9044
6087…
6088
Status.Temperature4 Float RO temperature of comp4
9045
6089…
6090
Status.Temperature5 Float RO temperature of comp5
9051
6101…
6102
Setup.Calibration.CurrentZerogas1 Float R/W current zero gas of comp1
9052
6103…
6104
Setup.Calibration.CurrentZerogas2 Float R/W current zero gas of comp2
9053
6105…
6106
Setup.Calibration.CurrentZerogas3 Float R/W current zero gas of comp3
9054
6107…
6108
Setup.Calibration.CurrentZerogas4 Float R/W current zero gas of comp4