Modbus Commands
Chapter 9
Modbus Functions
This chapter lists all Modbus functions
and registers supported by X-STREAM gas
analyzers.
Refer to the www.Modbus-IDA.org website
for detailled documentation about program-
ming the interface. At date of creation of this
instruc-tion manual the following documents
were used:
• MODBUS Protocol Specication:
Modbus_Application_Protocol_V1_1a.
pdf
• MODBUS Serial Line Implementation Guide:
Modbus_over_serial_line_V1.pdf.
9.1 Abstract
For a list of
supported functions page 9-2
supported parameters and registers,
ordered by parameter tag name page 9-22
ordered by register number page 9-20
Before using Modbus TCP/IP take care to
congure the communication properly:
6-79.
For Modbus TCP/IP the analyzer is factory
congured to support DHCP servers: The mo-
ment, the powered instrument is connected
9.1.1 Modbus TCP/IP
to a DHCP server via ethernet, it will receive
a valid IP address and become visible in the
network.
If no DHCP server is available, congure the
IP address manually
6-79.