X-STREAM XE
Instruction Manual
HASXEE-IM-HS
06/2010
6
Software Menus
Ch1
Calibration
Gases..
Tolerances..
Procedure..
Valve Assignment..
Program Sequence..
Interval Time..
Deviations..
6.2.3.2 Calibration Setup Menu
If the system is setup accordingly, the access
code for level 2 must be entered to gain ac-
cess to this menu.
6.2.3 Setup Menu
Submenu to specify the calibration gases for
the selected channel;
6-34.
Setup calibration tolerances;
6-35.
Setup details for the channel´s calibration
procedure;
6-36.
Assign calibration valves;
6-37.
Program a detailled calibration sequence;
6-39.
Specify interval times for automatic calibra-
tions;
6-40.
See (and reset) calibrations deviations infor-
mation
6-42.
Setup..
Calibration..
Note!
Multi-channel unit:
Some of the following submenus require to
select the component to be congured:
In SELECT COMPONENT, select which
channel´s calibration to be congured.