7.5 Replacing the Electrochemical Sensor
locking
device
7.5.2.3 How to Open X-STREAM XEF / XDF
To open a X-STREAM XEFD loosen the 20
screws located at the instrument´s ange.
Then carfeully ip down the front door to not
damage the instrument, hinges or equipment
installed below the analyzer.
7.5.2.4 How to Open X-STREAM XEFD
Depending on the individual analyzer con-
guration, either open the upper or lower
front door to the left, utilizing the two sash
fasteners.
Fig. 7-11: X-STREAM Field Housings - Interior Views
(shown Without Front Doors)
Note 1!
The internal box covering the physical components, as
shown in this gure, is optional and may not be installed in
your specic instrument!
Note 2!
To remove the box loosen the 2 screws of the locking device,
push it´s slider upwards as shown to the right, and take out
the box to the frontside of the instrument!
Locking device details
push here
to unlock
push this
way to lock
locked position unlocked position
EXPLOSION HAZARD
Special variations of X-STREAM XEF and XDF. as well as X-STREAM XEFD
may be installed in hazardous areas.
Maintaining such instruments is permitted only considering special
conditions, given in the related separate manuals.
Do not open nor maintain instruments in hazardous areas without having
read and understood all associated instruction manuals!
sash fasteners