Maintenance & Procedures
7.4.3 Advanced Calibration
The analyzer immediately begins span
calibration(s), showing the CALIBRATION
STATUS SUMMARY screen.
Press
enter in the rst line to cancel the cur-
rent calibration
"Current action" indicates, what currently is
carried out (purging, zeroing, ready)
"Action detail" shows the current procedure,
or Off
"Current duration" gives the remaining time
for the current procedure
"Prev. duration" shows the time elapsed since
start of procedure
"Current step" gives information about the
step currently carried out.
To see a detailled calibration status for a sin-
gle channel, enter CALIBRATION STATUS
SINGLE,
Calibration status summary
Cancel!
Calibration status single..
Current action Ready
Action detail Ch2
Current duration 0 s
Prev. duration 0 s
Current Step 0
Multi-channel unit:
Select the channel in SELECT COMPO-
NENT.
to open the status screen with enhanced
calibration information about the current
channel (indicated in the uppermost display
line), including
remaining calibration time, currently measured
concentration, setup of zero & span gas con-
centrations and currently calibrated range.
The procedure has nished when 'Applied gas'
shows Sample gas, or "Current action" in the
previous screen says Ready.
Press home to return to the MEASUREMENT
SCREEN.
Ch1
Calibration status single
Cancel!
Calibr.status Ready
Remaining time 0 s
Concentration 0.000 ppm
Zero gas 0.000 ppm
Span gas 5000.000 ppm
Current range Range 1
Applied gas Zero gas