Emerson HASXEE-IM-HS TV Video Accessories User Manual


 
Emerson Process Management GmbH & Co. OHG 8-5
X-STREAM XE
Instruction Manual
HASXEE-IM-HS
06/2010
8
Troubleshooting
Message
Status level
Explanation Recommended Actions
Factory le read error
Maintenance
Reading the factory congu-
ration le is not possible
• Check if internal disk is pre-
sent / installed
• Call Emerson service
Factory le write error
Maintenance
Writing the factory congura-
tion le is not possible. This
message does not appear
during normal operation!
FATAL!! Conguration
data destroyed
Failure
The instrument is now uncon-
gured, because retrieving
the conguration data from
several sources failed.
FATAL: FRAM read/write
error
Maintenance
The instrument is now uncon-
gured, because retrieving the
conguration data from sever-
al sources failed.
Call Emerson service
Flash write count over limit
Maintenance
Write cycles to internal CPU
Flash Memory exceeded
number of 90,000
Limitation analog output 1
Off-spec
Concentration assigned to
the indicated analog output
is outside congured ranges:
Analog output is limited to
congured ranges
• Use another measurement
range.
• Extend analog output range
conguration if possible.
• Run measurement inside its
given ranges.
Limitation analog output 2
Off-spec
Limitation analog output 3
Off-spec
Limitation analog output 4
Off-spec
Limitation analog output 5
Off-spec
PLC program error
Maintenance
While installing the PLC pro-
gram, a program error was
discovered
Check the PLC program for
syntax errors, wrong com-
mands or references
8.2.1 Analyzer Related Messages