Principles
3.6 Measurement Specication
Special Performance Specications for Gas Purity Measurements (ULCO & ULCO
2
)
Tab. 3-9: Special Performance Specications for Gas Purity Measurements
All these performance data is veried during the manufacturing process for each unit by the
following tests:
Linearization and sensitivity test
Long term drift stability test
Climate chamber test
Cross interference test (if applicable)
A61: 0 - 10 ... 100 ppm CO
A62: 0 - 20 ... 100 ppm CO
B62: 0 - 10 ... 100 ppm CO
2
B61: 0 - 5 ... 100 ppm CO
2
Noise < 1 %
1 2
< 1 %
1 2
resp. < 0.1 ppm
1 2 9
Detection limit <
2 %
1 2
<
2 %
1 2
resp. < 0.2 ppm
1 2 9
Linearity < +1 %
1 2
< +1 %
1 2
Zero-point drift < +2 %
1 2 3
< +2 %
1 2 3
resp. < 0.2 ppm
1 2 3 9
Span (sensitivity) drift < +2 %
1 2 4
< +2 %
1 2 4
resp. < 0.2 ppm
1 2 4 9
Repeatability < +2 %
1 2
< +2 %
1 2
resp. < 0.2 ppm
1 2 9
Response time (t
90
) < 10 s
7
< 10 s
7
Permissible gas ow 0.2 - 1.5 l/min. 0.2 - 1.5 l/min.
Inuence of gas ow < +2%
1 2
< +2%
1 2
Maximum gas pressure
10
<
1,500 hPa abs. (
<
7 psig)
<
1,500 hPa abs.
(
<
7 psig)
Inuence of pressure
– At constant temperature < 0.1 % per hPa
5
< 0.1 % per hPa
5
– With pressure compensation
8
< 0.01 % per hPa
5
<
0.01 % per hPa
5
Permissible ambient temperature 5 to +40 °C (41 to 104 °F) 5 to +40 °C (41 to 104 °F)
Inuence of temperature
(at constant pressure)
– On zero point
< +2 % per 10 K
6
< +2 % per 10 K
6
resp. < 0.2 ppm per 10 K
6 9
– On span (sensitivity) < +2 % per 10 K
6
< +2 % per 10 K
6
resp. < 0.2 ppm per 10 K
6 9
Thermostat control None None
1
Related to full scale
2
Constant pressure and temperature
3
Within 24 h; daily zero calibration requested
4
Within 24 h; daily span calibration recommended
5
Related to measuring value
6
Temperature must not change > 10 K within 1 h
7
From gas analyzer inlet at gas ow of 1.0 l/min
8
Barometric pressure sensor is required
9
Applies to measuring range < 10 ppm
10
Limited to atmospheric if internal sample pump