6.2.3 Setup Menu
6.2.3.6.6 Setup Alarms Calculatorn
Specify up to 4 threshold levels:
"LoLo" and "Lo" levels are to be setup below
the expected calculator result range.
"HiHi"and "Hi" levels are to be setup above
the calculator result range.
See the gure to the left for details.
Switch On or Off the alarm monitor for the
current calculator (here: exemplary Calculator
A).
Specify the hysteresis to be considered
around the threshold levels, to avoid oscilla-
ting alarms.
To view the current alarm states;
6-84
Any X-STREAM XE analyzer can be up-
graded with optional software packages to
provide 4 calculators (A ... D). The results can
be monitored to activate alarms.
Note!
This menu is available only, if a valid soft-
ware upgrade code has been purchased and
entered (
6-89 ).
For more information on calculators, see
the associated separate software features
manual.
Setup..
Alarms..
Calculator A..D
Calculator A
Alarm Monitor: On
LoLo Level: 50.000
Lo Level: 100.000
Hi Level: 400.000
HiHi Level: 600.000
Hysteresis: 1000
States..