Troubleshooting
8.2.1 Analyzer Related Messages
8.2.1 Analyzer Related NAMUR Messages
Message
Status level
Explanation Recommended Actions
Calculator program error
Maintenance
While running the calculator,
an inconsistency was detec-
ted
• Check the calculator pro-
gram for syntax errors, im-
possible commands or signal
references
• Check for divisions by 0
Division by 0 detected
Can´t open Data Logger
le
Maintenance
Accessing the data logger le
is not possible
• Check if internal disk is pre-
sent / installed
• Call Emerson Service
Can´t write to Data Logger
le
Maintenance
Accessing the data logger le
is not possible
Checksum error
Maintenance
Creating the factory congu-
ration le caused a checksum
error.
Store a new factory congura-
tion le.
Cfg checksum error
Maintenance
Creating the user congura-
tion le caused a checksum
error.
Create a new user congura-
tion le.
Cfg le open error
Maintenance
Opening the user congurati-
on le is not possible.
Call Emerson service
Cfg le read error
Maintenance
Writing to the user congurati-
on le is not possible.
Cfg le write error
Maintenance
Writing to the user congura-
tion le is not possible
Device not in Service
Function check
Operator has set the analyzer
to function check mode
Set analyzer into service
DISK Free space warning
Maintenance
Internal disk usage exceeded
specied limit (default: 80 %)
Free space by deleting les
DISK full
Maintenance
Internal disk usage exceeded
95 % limit