Using Spectrum Analyzer Features
Saving and Recalling Data from the Memory Card
To
Save
Limit-Line
l%bles
or
Amplitude
Correction
Factors
The procedure for saving limit-line tables or amplitude correction factors is
similar to saving trace data. State and trace data is not recalled when the
limit-line tables or amplitude correction factors are recalled.
1. Press
@iKKJ
or
@iK’Z],
Change Prefix to enter a new prelix or
change the existing prefix.
If you do not specify a prefix, the limit line table will be saved with a hle
name consisting of 1-(register number). A table of amplitude correction
factors will be saved with a
Ele
name consisting of
a-(register
number).
2. When saving tables, set up the table to be stored. See “Using the
Limit-Line Functions” or “Using Amplitude Correction Functions” for more
information about entering data.
3. Press
ISAVE_).
If INTERNAL is underlined, press INTERNAL CARD to select
CARD. Press Trace
--+
Card to access the menu with LIMIT LINES
and AMP
COR
4.
Press either LIMIT LINES , to save limit-line tables, or AMP COR , to save
amplitude-correction factors. REGISTER
#
and PREFIX= are displayed on
the cable TV analyzer display.
5. Use the numeric keypad to enter a register number and then press
(giEq.
The data is saved with a lile name consisting of a “1” (for limit-line tables)
or “a” (for amplitude-correction factors), the preEx that was entered, an
underscore (-), and the register number.