Spectrum Analyzer Mode Key Descriptions
Spectrum Analyzer Mode Functions
ANALYZER
ADDRESS
Option 021 only.
allows you to set the HP-IB address of the analyzer. The
analyzer address is set to 18 by pressing DEFAULT
CONFIG
.
Front-Panel Key Access
[??jXQ
ANALYZER
GAINS
displays the current value of the gain for various functional
blocks within the analyzer. The gain values will vary
depending on the current analyzer settings. This is a service
diagnostic function and is for service use only. Refer to the
service guide for more information.
Front-Panel Key Access (CAL)
ANNOTATN
ON OFF
screen annotation on or off. However, softkey annotation
will remain on the screen. The screen annotation may not
be required for prints or plots, or during remote operation.
Refer to the HOLD softkey description regarding softkey
annotation.
Front-Panel Key Access
cm)
APND
CAT ITEM
Refer to Chapter 4 of the HP 8590 E-Series and L-Series
S&W-urn
Analyzers and HP 8591 C Cable TV Analyzer
Programmer’s Guide for more information.
starts the DLP editor function and allows the highlighted
item from the catalog of analyzer memory to be appended
to the end of the item that is currently in the analyzer DLP
editor memory. If the item to be appended will not fit in the
available memory space, it will not be appended.
Front-Panel Key Access
[RECALL]
or
ISAVE)