Spectrum Analyzer Mode Key Descriptions
Spectrum Analyzer Mode Functions
SAVE
LIMIT
saves the current limit-line tables in the current
mass-storage device ( analyzer memory or memory card). To
verify the current mass-storage device, press SAVE LIMIT .
If MAX REG
#
appears on the analyzer display, the current
mass-storage device is analyzer memory. If PREFIX= is
displayed, the memory card is the mass-storage device.
Press (SAVE_) or
CRECALLI),
then INTERNAL CARD to change
the current mass-storage device. Press SAVE LIMIT, enter
a register number, then press
(ENTER)
to save the current
limit-line table in analyzer memory or on the memory card.
When saved on the memory card, limit-line tables are stored
with “1-“, the prefix, and the register number entered.
When saved in analyzer memory, the register number is
saved in a trace register. Trace-register values are restricted
to a range between 0 and the number x indicated by MAX
REG # = x.
Front-Panel Key Access
cm)
SCALE
LOG LIN
scales the vertical graticule divisions in logarithmic units
when LOG is underlined. When the SCALE LOG LIN
function is the active function, the logarithmic units per
division can be changed. Values may range from 0.1 to
20
dB
per division. When LIN is underlined, the vertical
scale is in linear mode which has a range of 1
kW
to 1
pW.
The reference-level value is set to the top of the screen and
the bottom graticule becomes zero volts. (Each division of
the graticule is one-eighth of the reference level in volts.)
Pressing SCALE LOG LIN always sets the units specified for
the current amplitude scale. Pressing
(-1
or powering
on the analyzer sets the default units.