HP (Hewlett-Packard) 8591C Cable Box User Manual


 
I
-
I
-
Making Measurements Using Spectrum Analyzer Mode
Stimulus-Response Measurements
4. Put the sweep time of the analyzer into stimulus-response auto-coupled
mode by pressing More
I
of 2 , then SWP CPLG
SR
SA until SR
(stimulus-response mode) is underlined. Auto-coupled sweep times are
usually much faster for swept-response measurements than they are for
spectrum analyzer measurements.
NOTE
In the stimulus-response mode, the II (reactance versus resistance) of the OUT can determine the
fastest rate at which the spectrum analyzer can be swept. To determine whether the analyzer is
sweeping too fast, slow the sweep time and note whether there is a frequency or amplitude shift of
the trace. Continue to slow the sweep time until there is no longer a frequency or amplitude shift.
I
4
REF
48.8
<
PEAK
/
/
1
lBmV AT
10
dB
1
SPAN
I500
0
MHz
u
I
\
FULL
SPQN
ZERO
SPAN
LAST
SPAN
:ENTER
350
.O
MHz
SPAN
500.0
MHz
#RES
BW
3.0
MHz VBW
1
MHz
SWP
20.0
msec
Figure 5-6. Spectrum Analyzer Settings According to the Measurement Requirement