Spectrum Analyzer Mode Key Descriptions
Spectrum Analyzer Mode Functions
INTERNAL
CARD
selects between spectrum analyzer memory and the memory
card for the save and recall functions.
Front-Panel Key Access
[RECALL_)
or
m
INTERNAL
+
STATE
recalls the saved analyzer state from the selected state
register. Recalling a state from the analyzer memory
displays the time and date when the state data was stored.
To recall a state, press INTERNAL
+
STATE and use the
numeric keypad to enter a state register number (valid state
register numbers are 1 through 9). State register 9 contains
a previous state; state register 0 contains the current state.
If windows are being used, the instrument state can only be
recalled into the active window.
Front-Panel Key Access
[RECALL_)
Internal
--f
Trace
accesses a softkey menu that allows you to either select the
trace in which the trace data is to be recalled (trace A, trace
B, or trace C), recall the current limit-line tables, or recall
amplitude correction factors. When recalling a trace, select
the trace in which the trace data is to be recalled, enter
the trace register number, and press
Cm).
If windows
are being used, only the trace of the active window can be
recalled.
When recalling limit-line tables or amplitude correction
factors, press LIMIT LINES or AMP
COR
respectively,
enter the trace register number, and press
CENTER).
Valid
trace register numbers are 0 through the maximum
register number. The maximum register number is the
number displayed after MAX REG
#
= during a save or recall
operation. If a screen title is present, it is recalled with
the trace data (but not with the limit-line table or the
amplitude-correction factors). If the screen title does not
exceed 34 characters, the time and date when the data was
stored will also be displayed. INVALID SAVEREG is displayed
if data has not been stored in the trace register.