I
-
CHANNEL
SPACING
Clear
Spectrum Analyzer Mode Key Descriptions
Spectrum Analyzer Mode Functions
allows the user to enter the channel spacing to set up the
analyzer when using the measurement functions under the
Power
Menu
softkey.
When the power measurements
are first accessed the initial value for channel spacing is
25
kHz.
Once the value is changed, the new value will be
saved through an instrument preset or power on. An error
message will occur for invalid values of channel spacing and
channel bandwidth.
If one of the power measurements is active and the channel
spacing is changed, with
PARAM
AUTO selected, then
the coupled analyzer settings are immediately updated. If
CONT
MEAS
is also selected, then another sweep is taken
and the measured results are updated.
Front-Panel Key Access
(MEAS/USERj
clears the current screen title or orefix. This softkey is
accessed under the
YZ,%
Spc Clear softkey when using
Change Title or Change Prefix.
Front-Panel Key Access (CAL),
[CONFIGI),
(jjl,
(ml,
or (SAVE)
CLEAR
PARAM
clears all of the pulse parameters in the gate utility. It sets
the value of pulse width, pulse repetition interval, and
reference edge to zero and turns off the coupling of the pulse
parameters to the resolution bandwidth, video bandwidth,
and sweep time.
Front-Panel Key Access [SWEEP]