Percent amplitude modulation can be measured quickly and easily using
the one-button % AM function. The signal and both its sidebands must be
on the display. The sidebands are assumed to be entirely from amplitude
modulation. The spectrum analyzer places arrow markers on the three signals
to be used to compute percent amplitude modulation, and displays the value.
If the sidebands are not in the frequency span or their frequency spacing is
not equal, the measurement stops and an error message is displayed.
Percent AM measurement accuracy for close signals:
l typically about
f0.
1
%/%
, for log mode
l typically about
*3
% , for linear mode