Spectrum Analyzer Mode Key Descriptions
Spectrum Analyzer Mode Functions
SAVE
AMP
CQR
saves the current amplitude-correction factors table to
the current mass-storage device ( analyzer memory or
memory card). To verify the current mass storage device,
press SAVE AMP COR If MAX REG
#
appears on the
analyzer display, the current mass storage device is analyzer
memory. If PREFIX= is displayed, the memory card is
the mass storage device. Press
ISAVE)
or
cm),
then
INTERNAL CARD to change the current mass storage device.
Press SAVE AMP COR , enter a register number, then press
@KiK)
to save the current amplitude-correction factors
table in analyzer memory or on the memory card. When
saved on the memory card, amplitude-correction factors
tables are stored with “a-“, the prefix, and the register
number entered. When saved in analyzer memory, the
register number is saved in a trace register. Trace register
values are restricted to a range between 0 and the number x
indicated by MAX REG#
=
x.
SAVE
EDIT
Front-Panel Key Access
ICAL)
Refer to the HP 8590 E-Series and
L-Serks
Spectrum
Analgws
and HP 8591 C Cable TV Analyzer Programmer’s
Guide for more information.
passes the text from the DLP editor memory through
the parser to execute as analyzer commands. Pressing
SAVE
EDIT
is similar to outputting the text to the analyzer
from an external controller. If the text (commands) is a valid
user-defined function, it passes through the parser and into
the analyzer user memory. It will replace an existing user
defined function of the same name.
The DLP editor memory buffer remains intact when the
analyzer is preset or powered off so the text being edited
will not be lost.