Error Messages
Marker Count Widen RES BW
Indicates that the current resolution bandwidth setting is too narrow to use
with the marker counter function. The marker counter function can be in
narrow resolution bandwidths (bandwidths that are less than 1
kHz)
with
the following procedure:
1. Place the marker on the desired signal.
2. Increase the resolution bandwidth to 1
kHz
and verify the marker is on
the signal peak.
3. If the marker in on the signal peak, the marker count function can be
used in either the 1
kHz
resolution bandwidth or the original narrow
resolution bandwidth setting. If the marker is not on the signal peak,
it should be moved to the signal peak and the marker counter function
should not be used with a resolution bandwidth setting of less than
1
kHz.
(U>
MEAS
UNCAL
The measurement is uncalibrated. Check the sweep time, span, and
bandwidth settings, or press [AUTO COUPLE), AUTO ALL (U)
MEMORY
0VERFLOW:ERASE
DLP MEM and reload
This message indicates that too many user-defined items (functions,
variables, key deEnitions), or downloadable programs have been
loaded into cable TV analyzer memory. If this message appears, use
ERASE DLP
MEN
and then load the user-deEned item or downloadable
program into cable TV analyzer memory. (U)
No card found
Indicates that the memory card is not inserted. (U)
No points defined
Indicates the speciEed limit line or amplitude correction function cannot be
performed because no limit line segments or amplitude correction factors
have been deEned. (U)
OVEN COLD
Indicates that the cable TV analyzer has been powered up for less than 5
minutes. (The actual temperature of the precision frequency oven is not
measured.) (M)