Preparing For Use
Turning on the Analyzer for the First Time
7. When the frequency and amplitude self-calibration routines have
been completed successfully, store the correction factors by pressing
CAL STORE.
The self-calibration routines calibrate the cable TV analyzer by generating
correction factors. The softkey CAL STOKE stores the correction factors
in the area of cable TV analyzer memory that is stored when the cable TV
analyzer is turned off; the cable TV analyzer will automatically apply these
factors in future measurements. If CAL STOKE is not pressed, the correction
factors remain in effect until the cable TV analyzer is turned off.
Performing
the
Tracking-Generator
Self-Calibration
Routine
For cable TV analyzers with Option 011, the tracking-generator
self-calibration routine should be performed prior to using the tracking
generator.
NOTE
Since the tracking generator calibration routine depends on the accuracy of the absolute amplitude level
of the cable TV analyzer, the cable TV analyzer amplitude calibration should be done prior to using
CAL TRK
GEN