Spectrum Analyzer Mode Key Descriptions
Spectrum Analyzer Mode Functions
ALC
PITR
INT XTAL
Option 011 only.
activates the automatic leveling control (ALC) function for
internal
(INT)
leveling or external (XTAL or MTR) leveling.
The external leveling input (located on the rear panel of the
analyzer) can be used with a power meter or crystal that has
a positive or negative voltage output. See specifications
and characteristics in your calibration guide for the leveling
input characteristics. External leveling increases the
amplitude accuracy by improving the effective source match.
The meter (MTR) position narrows ALC loop bandwidth so
an HP power meter can be used.
Front-Panel Key Access
[AUXCTRL_)
ALC TEST
Option 011 only
breaks the leveling loop of the automatic leveling control in
the tracking generator. This is a service diagnostic function
and is for service use only. Refer to the service guide for
more information.
Front-Panel Key Access
ICALl
ALL DLP
--t CARD
saves
all
the downloadable programs and key definitions
that are in analyzer memory onto the memory card. If
the downloadable program was stored using a prefix,
the file name for the downloadable program consists of
d(prefix)-(register
number). If no prefix was specified, the
data is stored with the file name
d-(register
number).
Front-Panel Key Access
m
Amp Cor
accessed by
ICAL).
Amp Cbr accesses the menus for
entering and editing the current amplitude-correction
factors.
Front-Panel Key Access (CAL)