Making Basic Measurements Using Spectrum Analyzer Mode
Tracking Unstable Signals Using Marker Track and the Maximum Hold and Minimum Hold Functions
5. The signal frequency drift can be read from the screen if both the marker
track and marker delta functions are active. Press
(MKR),
MARKER
A
,
I’jMKRFCTN),
MK
TRACK
ON OFF
;
th
e marker readout indicates the change
in frequency and amplitude as the signal drifts. (See Figure 4-8.)
MK
NOISE
ON
OFF
.ENTER
100
0958
MHZ
SPAN
500.0
KHZ
XRES
BW
10
kHZ
VBW
10
kHZ
SWP
30.0
mset
Figure 4-8. Using Marker Tracking to Track an Unstable Signal
The spectrum analyzer can measure the short- and long-term stability of a
source. The maximum amplitude level and the frequency drift of an input
signal trace can be displayed and held by using the maximum-hold function.
The minimum amplitude level can be displayed by using minimum hold
(available for trace C only).
You can use the maximum-hold and minimum-hold functions if, for example,
you want to determine how much of the frequency spectrum an FM signal
occupies.