HP (Hewlett-Packard) 8591C Cable Box User Manual


 
I
-
I
-
If You Have a Problem
Error Messages
Gate card not calibrated
This message can indicate that either the CAL AMPTD routine need
to be performed before the time-gating functions can be used, or that
something was connected to the GATE TRIGGER INPUT connector during
the CAL AMPTD or CAL FREQ
&
AMPTD routines. Ensure that nothing is
connected to the GATE TRIGGER INPUT connector when the CAL AMPTD
or CAL FREQ
&
AMPTD routines are performed. (U) and (H)
INTERNAL LOCKED
The cable TV analyzer internal trace and state registers have been locked.
To unlock the trace or state registers, press SAV LOCK ON OFF so that
OFF is underlined. For remote operation, use PSTATE OFF. (U)
INVALID ACTDEF:
_
-
-
The specified ACTDEF name is not valid. See the ACTDEF programming
command. (U)
INVALID AMPCOR: FREQ
For the AMPCOR command, the frequency data must be entered in
increasing order. See the description for the AMPCOR programming
command for more information. (U)
INVALID BLOCK FORMAT: IF STATEMENT
An invalid block format appeared within the IF statement. See the
description for the IF THEN ELSE
ENDIF
programming command for more
information. (U)
INVALID CARD
Indicates one of the following conditions: a card reader is not installed, the
memory card is write-protected (check the position of the switch on the
memory card), the memory card is a read-only memory (ROM) card, or a
memory card has not been inserted.
INVALID CARD: BAD MEDIA
Indicates the formatting routine (FORMAT CARD
)
for the memory card
could not be completed. See the description for INVALID CARD above for
more information about the possible causes of this message. (U) and (H)
INVALID CARD: DATA ERROR
Indicates the data could not be retrieved from the memory card. (U) and
(HI
8-19
I-