Teledyne 9110T TV Converter Box User Manual


 
Model 9110TH NOx Analyzer Setup Menu
Teledyne Analytical Instruments 125
5.9.3.7. Manual Adjustment of Current Loop Output Span and Offset
A current loop option may be purchased for the A1, A2 and A3 Analog outputs of the analyzer. This
option places circuitry in series with the output of the D-to-A converter on the motherboard that changes
the normal DC voltage output to a 0-20 milliamp signal (See Section 3.3.1.4).
1. The outputs can be ordered scaled to any set of limits within that 0-20 mA range,
however most current loop applications call for either 0-20 mA or 4-20 mA range
spans.
2. All current loop outputs have a +5% over range. Ranges whose lower limit is set above 1 mA also have
a –5% under range.
To switch an analog output from voltage to current loop, follow the instructions in Section 5.9.3.1 (select
CURR from the list of options on the “Output Range” menu).
Adjusting the signal zero and span levels of the current loop output is done by raising or lowering the
voltage output of the D-to-A converter circuitry on the analyzer’s motherboard. This raises or lowers the
signal level produced by the current loop option circuitry.
The software allows this adjustment to be made in 100, 10 or 1 count increments. Since the exact
amount by which the current signal is changed per D-to-A count varies from output-to-output and
instrument–to–instrument, you will need to measure the change in the signal levels with a separate,
current meter placed in series with the output circuit. See Figure 3-8 for pin assignments and diagram of
the analog output connector.
Figure 5-7: Setup for Checking / Calibration Current Output Signal Levels Using an Ammeter
CAUTION GENERAL SAFETY HAZARD
Do not exceed 60 V peak voltage between current loop outputs and instrument ground.