Model 9110TH NOx Analyzer Getting Started
Teledyne Analytical Instruments 79
Cal gas will be supplied through the SAMPLE gas inlet on the back of the analyzer and;
The pneumatic setup matches that described in Section 3.3.2.
VERIFYING THE REPORTING RANGE SETTINGS
While it is possible to perform the following procedure with any range setting we recommend that you
perform this initial checkout using following reporting range settings:
Unit of Measure: PPB
Reporting Range: 500 ppb
Mode Setting: SNGL
While these are the default setting for the 9110T analyzer, it is recommended that you verify them before
proceeding with the calibration procedure, by pressing the following menu button sequence: