Teledyne 9110T TV Converter Box User Manual


 
Model 9110TH NOx Analyzer Getting Started
Teledyne Analytical Instruments 31
3. GETTING STARTED
This section addresses the procedures for unpacking the instrument and inspecting for damage, presents
clearance specifications for proper ventilation, introduces the instrument layout, then presents the
procedures for getting started: making electrical and pneumatic connections, and conducting an initial
calibration check.
3.1. UNPACKING THE 9110T ANALYZER
CAUTION
GENERAL SAFETY HAZARD
To avoid personal injury, always use two persons to lift and carry the 9110T.
ATTENTION
COULD DAMAGE INSTRUMENT AND VOID WARRANTY
Printed Circuit Assemblies (PCAs) are sensitive to electro-static
discharges too small to be felt by the human nervous system. Failure to
use ESD protection when working with electronic assemblies will void
the instrument warranty. Refer to Section 14 for more information on
preventing ESD damage.
CAUTION!
Do not operate this instrument until you’ve removed dust plugs from SAMPLE
and EXHAUST ports on the rear panel.
Note TAI recommends that you store shipping containers/materials for future
use if/when the instrument should be returned to the factory for repair
and/or calibration service. See Warranty section in this manual and
shipping procedures on our Website at http://www.teledyne-api.com
under Customer Support > Return Authorization.
Verify that there is no apparent external shipping damage. If damage has occurred, please advise the
shipper first, then TAI.
Included with your analyzer is a printed record of the final performance characterization performed on
your instrument at the factory. This record, titled Final Test and Validation Data Sheet (P/N 04490) is
an important quality assurance and calibration record for this instrument. It should be placed in the
quality records file for this instrument.