Model 9110TH NOx Analyzer Communications Setup and Operation
Teledyne Analytical Instruments 143
Table 6-3. LAN/Ethernet Default Configuration Properties
PROPERTY DEFAULT STATE DESCRIPTION
DHCP
ON
This displays whether the DHCP is turned ON or OFF. Press
EDIT and toggle ON for automatic configuration after first
consulting network administrator.
INSTRUMENT
IP ADDRESS
This string of four packets of 1 to 3 numbers each (e.g.
192.168.76.55.) is the address of the analyzer itself.
GATEWAY IP
ADDRESS
0.0.0.0
Can only be edited when DHCP is set to OFF.
A string of numbers very similar to the Instrument IP address
(e.g. 192.168.76.1.) that is the address of the computer used by
your LAN to access the Internet.
SUBNET MASK
0.0.0.0
Can only be edited when DHCP is set to OFF.
Also a string of four packets of 1 to 3 numbers each (e.g.
255.255.252.0) that identifies the LAN to which the device is
connected.
All addressable devices and computers on a LAN must have the
same subnet mask. Any transmissions sent to devices with
different subnets are assumed to be outside of the LAN and are
routed through the gateway computer onto the Internet.
TCP PORT
1
3000
This number defines the terminal control port by which the
instrument is addressed by terminal emulation software, such as
Internet or Teledyne API’s APICOM.
HOST NAME
[initially blank]
The name by which your analyzer will appear when addressed
from other computers on the LAN or via the Internet. To assign
or change, see Section 6.5.2.1.
1
Do not change the setting for this property unless instructed to by TAI’s Customer Service personnel.
6.5.2. CONFIGURING ETHERNET COMMUNICATION USING DYNAMIC
HOST CONFIGURATION PROTOCOL (DHCP)
The default Ethernet setting is DHCP.