Model 9110T NOx Analyzer Troubleshooting & Service
Teledyne Analytical Instruments 294
Input the starting NO
2
concentration value into the 9110T by pressing:
Converter Efficiency Menu
SAMPLE RANGE=500.0 PPB NO=XXXX
<TST TST> CAL SETUP
SAMPLE RANGE TO CAL
LOW HIGH ENTR EXIT
M-P CAL CONCENTRATON MENU
NOX NO CONV EXIT
M-P CAL CONVERTER EFICIENCY MENU
NO2 CAL SET EXIT
M-P CAL NO2 CE CONC: 500.0 Conc
0400.00ENTREXIT
SAMPLE RANGE=500.0 PPB NO=XXXX
<TST TST> CAL SETUP
This step only appears if the
analyzer’s reporting range is
set for AUTO range mode.
Select LOW and press ENTR.
Repeat entire procedure for
HIGH range.
The expected NO
2
span
concentration value defaults
to 400.0 Conc.
Make sure that you specify
the actual concentration value
of the NO
2
gas.
Toggle these buttons
to change this value to
the concentration of
the NO
2
gas being
used.
Use these buttons to
select the appropriate
range.
Repeat entire procedure
for each range.
SAMPLE RANGE=500.0 PPB NO=XXXX
<TST TST> ZERO CONC SETUP