Model 9110T NOx Analyzer Data Acquisition System (DAS) and APICOM
Teledyne Analytical Instruments 166
These default data channels can be used as they are, or they can be customized from the front panel to fit
a specific application. They can also be deleted to make room for custom user-programmed Data
Channels.
Appendix A lists the firmware-specific DAS configuration in plain-text format. This text file can either
be loaded into APICOM and then modified and uploaded to the instrument or can be copied and pasted
into a terminal program to be sent to the analyzer.
IMPORTANT
IMPACT ON READINGS OR DATA
Sending a DAS configuration to the analyzer through its COM ports will
replace the existing configuration and will delete all stored data. Back up
any existing data and the DAS configuration before uploading new
settings.