Teledyne 9110T TV Converter Box User Manual


 
Model 9110T NOx Analyzer Calibration Procedures
Teledyne Analytical Instruments 218
EXIT returns
to the SETUP
Menu.
SETUP DURATION:15.0 MINUTES
<SET SET> EDIT EXIT
SETUP DURATION 15.0MINUTES
3 0 .0 ENTR EXIT
SETUP DURATION:30.0 MINUTES
<SET SET> EDIT EXIT
Toggle buttons to set
duration for each iteration
of the sequence:
Set in Decimal minutes
from 0.1 – 60.0.
SETUP CALIBRATE: OFF
<SET SET> EDIT EXIT
SETUP CALIBRATE: OFF
ON ENTR EXIT
SETUP X.X CALIBRATE: ON
<SET SET> EDIT EXIT
Toggle button
Between Off and
ON.
SETUP X.X SEQ 2) ZERO
SPAN, 2:00:30
PREV NEXT MODE SET EXIT
Display show:
SEQ 2) ZERO–SPAN, 2:00:30
Sequence Delta Time
MODE Delta Days
CONTINUED FROM PREVIOUS PAGE
DELTA TIME
Note If at any time an unallowable entry is selected (Example: Delta Days > 367)
the ENTR button will disappear from the display.