Model 9110TH NOx Analyzer Communications Setup and Operation
Teledyne Analytical Instruments 147
6.6. USB PORT FOR REMOTE ACCESS
The analyzer can be operated through a personal computer by downloading the TAPI USB driver and
directly connecting their respective USB ports.
1. Install the Teledyne T-Series USB driver on your computer, downloadable from the Teledyne API
website under Help Center>Software Downloads (www.teledyne-api.com/software).
2. Run the installer file: “TAPIVCPInstaller.exe”
3. Connect the USB cable between the USB ports on your personal computer and your analyzer. The
USB cable should be a Type A – Type B cable, commonly used as a USB printer cable.
4. Determine the Windows XP Com Port number that was automatically assigned to the USB
connection. (Start → Control Panel → System → Hardware → Device Manager). This is the com
port that should be set in the communications software, such as APIcom or Hyperterminal.
Refer to the Quick Start (Direct Cable Connection) section of the Teledyne APIcom Manual, PN 07463.
5. In the instrument’s SETUP>MORE>COMM>COM2 menu, make the following settings: