Model 9110TH NOx Analyzer Specifications, Approvals, & Compliance
Teledyne Analytical Instruments 27
2. SPECIFICATIONS, APPROVALS, & COMPLIANCE
This section presents specifications for the 9110T, Agency approvals, EPA designation, and CE mark
and safety compliance..
2.1. SPECIFICATIONS
Table 2-1 presents the instrument’s parameters and the specifications that each meets.
Table 2-1: 9110T Basic Unit Specifications
Parameter
Specification
Min/Max Range
(Physical Analog Output)
Min: 0-50 ppb Full Scale
Max: 0-20,000 ppb Full Scale (selectable, independent NO, NO
2
, NO
x
ranges and
auto ranges supported)
Measurement Units ppb, ppm, µg/m
3
, mg/m
3
(selectable)
Zero Noise
1
< 0.2 ppb (RMS)
Span Noise
1
< 0.5% of reading (RMS) above 50 ppb
or 0.2 ppb, whichever is greater
Lower Detectable Limit
2
0.4 ppb
Zero Drift < 0.5 ppb (at constant temperature and voltage) /24 hours
Span Drift < 0.5% of Full Scale (at constant temperature and voltage) /24 hours
Lag Time
1
20 seconds
Rise/Fall Time
1
<60 seconds to 95%
Linearity 1% of full scale / 24 hours
Precision 0.5% of reading above 50 ppb
Sample Flow Rate 500 cm
3
/min ± 10%
AC Power 100-120 V, 60 Hz (3.0 A); 220-240 V, 50 Hz (2.5 A)
Power, Ext Pump
100 V, 50/60 Hz (3.25 A); 115 V, 60 Hz (3.0 A);
220-240 V, 50/60 Hz (2.5 A)
Analog Output Ranges
10V, 5V, 1V, 0.1V (selectable)
All Ranges with 5% Under/Over Range
Analog Output Resolution 1 part in 4096 of selected full-scale voltage
Recorder Offset
± 10%
Standard I/O 1 Ethernet: 10/100Base-T
2 RS-232 (300 – 115,200 baud)
2 USB device ports
8 opto-isolated digital status outputs (7 defined, 1 spare)
6 opto-isolated digital control inputs (4 defined, 2 spar)
4 analog outputs
Optional I/O 1 USB com port
1 RS485
8 analog inputs (0-10V, 12-bit)
4 digital alarm outputs
Multidrop RS232
3 4-20mA current outputs
Dimensions H x W x D 7" x 17" x 23.5" (178mm x 432 mm x 597 mm)
Weight Analyzer: 40 lbs (18 kg)