Teledyne 9110T TV Converter Box User Manual


 
Model 9110T NOx Analyzer Setup Menu
Teledyne Analytical Instruments 98
reset the upper range limits for the new mode selection..
5.4.2.2. Analog Output Reporting Range Default Settings
The default setting for these the reporting ranges of the analog output channels A1, A2 and A3 are:
SNGL mode
0 to 500.0 ppb
0 to 5 VDC
5.4.3. SETUP RNGE MODE
Single range mode (SNGL) reports all three of the NOx gas concentrations using the same reporting
range span (see Section 5.4.3.1).
Independent range mode (IND) allows the NOx, NO and NO2 analog outputs to be set with different
reporting range spans (see Section 5.4.3.2).
Automatic range mode (AUTO) allows the analyzer to automatically switch the reporting range between
two user upper span limits (designated LOW and HIGH) based on the actual concentrations being
measured for each (see Section 5.4.3.3). These are not the same as the analyzer’s low and high physical
ranges.
5.4.3.1. SETUP RNGE MODE SNGL: Configuring the 9110T Analyzer for Single
Range Mode
Note
Single Range is the default reporting range mode for the analyzer.
When the single range mode is selected (SNGL), all analog NO
x
, NO and NO
2
concentration outputs
(A1, A2 and A3) are slaved together and set to the same reporting range limits (e.g. 500.0 ppb). This
reporting range can be set to any value between 100 ppb and 20,000 ppb.
Although all three NO
x
outputs share the same concentration reporting range, the electronic signal ranges
of the analog outputs may still be configured for different values (e.g. 0-5 VDC, 0-10 VDC, etc; see
Section 5.9.3.1).
To select SNGL range mode and to set the upper limit of the range, press: