To assign or reset the status flag bit assignments, press:
Concentration field
displays all gases.
SAMPLE RANGE=500.0 PPB NOX= XXXX
<TST TST> CAL SETUP
Press the [?] button repeatedly to cycle
through the available character set: 0-9
NOTE: Values of A-F can also be set
but are meaningless.
The <CH and CH>
buttons move the
cursor brackets “[ ]”
left and right along the
bit string.
SETUP X.X HESSEN STATUS FLAGS
<SET SET> EDIT EXIT
SETUP X.X COMMUNICATIONS MENU
ID HESN COM1 COM2 EXIT
SETUP X.X HESSEN VARIATION:TYPE1
<SET SET> EDIT EXIT
SETUP X.X PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK MORE EXIT
SETUP X.X SECONDARY SETUP MENU
COMM VARS DIAG EXIT
Continue pressing SET until ...
SETUP X.X IZS TEMP WARNING:0020
PREV NEXT EDIT PRNT EXIT
SETUP X.X BOX TEMP WARNING:[0]008
<CH CH> DEL [0] ENTR EXIT
SETUP X.X BOX TEMP WARNING:0008
PREV NEXT EDIT PRNT EXIT
Continue pressing NEXT until desired
flag message is displayed
EXIT discards the
new setting.
ENTR accepts the
new setting.
INS Inserts a the
character at the
current location of the
cursor brackets.
6.7.2.7. Instrument ID
Each instrument on a Hessen Protocol network must have a unique identifier (ID number). Refer to
Section 5.7.1 for information and to customize the ID of each.